B&K Precision 866C Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Medição, teste e controlo B&K Precision 866C. B&K Precision 866C User manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 76
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
USER MANUAL
8500 SERIES DC ELECTRONIC LOADS
Models 8500, 8502, 8510, 8512, 8514, 8518, 8520,
8522, 8524 & 8526
Vista de página 0
1 2 3 4 5 6 ... 75 76

Resumo do Conteúdo

Página 1 - 8522, 8524 & 8526

USER MANUAL8500 SERIES DC ELECTRONIC LOADSModels 8500, 8502, 8510, 8512, 8514, 8518, 8520, 8522, 8524 & 8526

Página 2 - Safety Symbols

Shift + Menu Enter the instrument's menu system.Shift + Short Turn short circuit on or off.Shift + Tran Start or stop transient condition.Shift +

Página 3 - Table of Contents

not covered by warranty.4Line voltage selection switch (110 VAC or 220 VAC)Display annunciatorsAnnunciator MeaningOFF The load is off.UNREG The input

Página 4

Lvl Menu item Function2 INITIAL CONFIG Return instrument to factory default settings.2 POWER-ON RECALL Power on state of instrument.3 ON Remembers sta

Página 5 - Notation

Lvl Menu item Function2 BAUDRATE SET Set the serial bus connection speed.348003 96003 192003 384002 COMM. PARITY SET Serial connection parity.3NONE3 E

Página 6 - Quick reference

Lvl Menu item Function4 REPEAT List is repeatedly executed after the trigger is received.3 VOLTAGE LIST4 ONCE List is executed once per trigger.4 REPE

Página 7 - Front panel

SpecificationsModels 8500 & 8502 (300 W)Parameter 8500 8502Input rating Voltage 0 to 120 V 0 to 500 V Current 1 mA to 30 A 1 mA to 15 A

Página 8

Models 8520/8522/8524/8526 (2400W & 5000W)Parameter 8520 8522 8524 8526Input rating Voltage 0 – 120 V 0 – 500 V 0 – 60 V 0 – 500 VCurrent 0 – 240

Página 9 - Front panel keys

Parameter SpecificationHumidity ≤ 95% relative humidity, non-condensingAltitude ≤ 2000 mAC Line voltage220 AV±10%,47~63 Hz110 AV±10%,47~63 HzOperating

Página 10 - Rear panel

The curved portion is where the dissipated power is at the rated power of the instrument (and is actually a hyperbolic shape). When you use the menu

Página 11 - Display annunciators

Slew rateThe slew rate for each DC load varies from model to model. The variations are also dependent on the different regions measured for each indi

Página 12

SafetyThe following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precau

Página 13

As a reference, below is a table of slew rates for their respective models:Model Slew rate85000.5A / S85020.5A / S85101A/ S85120.5A / S85141A/ S8

Página 14

Glossary△ Up arrow key. Used to scroll through the menu or cause the temporary display of the alternate standard display.▽ Down arrow key. Used to s

Página 15 - Specifications

R-set Configure the instrument for constant resistance mode.Recall Recall instrument state from non-volatile memory.Remote sensing Allows the instrume

Página 16 - Environmental Conditions

InstallationInspectionItems you should have receivedWhen you open the box containing the instrument, you should find the following items:1. The DC Loa

Página 17

Models 8510, 8512, 8514 & 8518Models 8520, 85228500 DC Load Series Version: November 4, 2008 Page 24 of 76

Página 18 - Low Voltage Characteristics

Models 8524 & 8526Unit (mm)8500 DC Load Series Version: November 4, 2008 Page 25 of 76

Página 19 - Slew rate

Model number Dimensions in mm Mass in kg8500 215W×88H×355D 5.28502 215W×88H×355D 5.28510 429W×88H×355D 148512 429W×88H×355D 148514 429W×88H×355D 14851

Página 20 - 0.5A / S

First turn-on checkoutCAUTIONEnsure that the line voltage selector switch on the back panel is set to match your line voltage. Failure to do so could

Página 21 - Glossary

watt.Press the R-set button (you may have to press it twice). Set the resistance to 100 Ω. Press the On/Off key. Verify that the current is approx

Página 22

Power-on stateThe default power-on instrument state is to not remember the instrument mode settings before the last power-down. If you wish to have t

Página 23 - Models: 8500& 8502

Table of ContentsSafety...

Página 24 - Models 8520, 8522

Key DisplayI-set If you press this key after powering up and there is no currently-stored constant current value, you'll be prompted for the desi

Página 25 - Models 8524 & 8526

Key Displaypower level. If there was a power value already stored, it will be momentarily displayed and the instrument will be in CW mode. If you wi

Página 26 - Bench operation

Key DisplayEnter :TIMER STATEEnter Use arrow keys to set to :ON.▽ :TIMER SETEnter TIMER=XXXXXS Enter desired time interval. Valid values are 1 to 60

Página 27 - First turn-on checkout

the battery) in ampere*hours (A*hrs) of the battery is calculated and displayed. To run a battery test, follow these steps:Key presses DisplayI-set,

Página 28 - Operational introduction

Keys DisplayShift + S-Tran LEVEL A= X.XXXA You are being prompted for the first current value. Press the 5 key for 5 A.Enter Enters the 5 A value, th

Página 29 - Constant current mode

Keys Displaycurrent value. Press the 1 and 0 keys to enter 10 A.Enter Enters the 10 A value, then displays WIDTH B = X.XMS. You're being promp

Página 30 - Constant power mode

Keys DisplayEnter Displays LEVEL B= X.XXXA. You are being prompted for the second current value. Press the 1 and 0 keys to enter 10 A.Enter Enters t

Página 31 - Timed operation

Current, A Duration, ms Between times3 1000 0 and 10 800 1 and 22 500 2 and 30 300 3 and 46 500 4 and 5This list has 5 steps, located at the transitio

Página 32 - Battery test

modes, mode parameters, and durations. They are useful for executing a set of tests on a device, then displaying whether the tests passed or failed.

Página 33 - 2.0ms 3.0ms

Keys Display2.5 Enter SHORT OFF We select short on and then press Enter. Enter▽ READBACK A We want to readback the current, so we press Enter.Enter

Página 35 - Toggled Transient Operation

Trigger Type Explanationfront panel keys.EXTERNAL An external trigger is a TTL high signal applied to the Trigger connection on the back panel. This

Página 36 - List sequence

front panel using the I-set, V-set, P-set, or R-set keys.To set these protection values, use the following keystrokes:Keys DisplayShift + Menu :CONFIG

Página 37 - Test files

reversed polarity. If a reverse voltage condition is detected, the buzzer will sound and REVERSE VOLTAGE will be displayed on screen.Over Temperature

Página 38

Example: A power supply is connected to the DC load with 72.5 cm of 20-gauge solid copper wires. The constant current i set to 5 A. The power supply

Página 39 - Triggering

Remote operationCommunication cablesThe DC Load has a DB9 connector on the rear panel that allows remote communication.WARNINGDo not connect the DC Lo

Página 40 - Protection features

This COM port can then be accessed as if it were a regular RS-232 port.The LEDs in the adapter will blink when information is being sent through the a

Página 41 - Reverse voltage protection

The checksum number is the arithmetic sum of each of the bytes modulo 256.Status packetsWhen you send a command that does not cause the DC Load to sen

Página 42 - Remote sensing

print " ", s = "%02x" % ord(bytes[i]) if s == "00": s = chr(250)*2 print s,

Página 43 - Saving and Recalling settings

Here are the printed results when the above script is ran:Set to remote command: aa ·· 20 01 ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·

Página 44 - INSTRUMENT

example programs. The Summary of commands section is a list of the commands, but without details. The Command details section explains how to use ea

Página 45 - RS-232 settings

NotationI-set Represents a key on the front panel. Note that some of these may be accessed in combination with the Shift key.OFF Annunciator, value,

Página 46 - Example program

Command Group Byte ActionList operations0x3A Select the list operation (CC/CV/CW/CR)0x3B Read the list operation (CC/CV/CW/CR)0x3C Set how lists repea

Página 47

Command Group Byte Action0x5C Recall DC Load's settingsFunction0x5D Select FIXED/SHORT/TRAN/LIST/BATTERY function0x5E Get function type (FIXED/SH

Página 48 - Chapter organization

Least significant two bytes, most significant byte Lower high byteMost significant two bytes, least significant byte Higher low byteMost significant t

Página 49 - Summary of commands

Byte offsetMeaning4 Lower high byte of maximum voltage.5 Upper low byte of maximum voltage.6 Upper high byte of maximum voltage.7-24 ReservedExample:

Página 50

Byte offsetMeaning5 Upper low byte of maximum current.6 Upper high byte of maximum current.7-24 Reserved0x26 Set the maximum power allowedByte offse

Página 51 - Notation for tables

Byte offsetMeaning3 Mode:0 is CC1 is CV2 is CW3 is CR4-24 Reserved0x2A Set CC mode current Byte offsetMeaning3 Lower low byte of current. 1 represe

Página 52

Byte offsetMeaning4 Lower high byte of voltage.5 Upper low byte of voltage.6 Upper high byte of voltage.7-24 Reserved0x2E Set CW mode powerByte offs

Página 53

Byte offsetMeaning5 Upper low byte of resistance.6 Upper high byte of resistance.7-24 Reserved0x32 Set CC mode transient current and timing Byte off

Página 54

Byte offsetMeaning0 is CONTINUOUS1 is PULSE2 is TOGGLED16-24 Reserved0x35 Read CV mode transient parametersByte offsetMeaning3 to 6 Value A of volta

Página 55 - 0x2D Read CV mode voltage

Byte offsetMeaning15 Transient operation:0 is CONTINUOUS1 is PULSE2 is TOGGLED16-24 Reserved0x38 Set CR mode transient resistance and timingByte off

Página 56 - 0x30 Set CR mode resistance

Quick referenceModel numbers covered by this documentThe B&K Precision DC Loads covered by this manual are:● 8500● 8502● 8510● 8512● 8514● 8518●

Página 57

Byte offsetMeaning3 is constant resistance (CR)4-24 Reserved0x3B Read the list operation (CC/CV/CW/CR)Byte offsetMeaning3 List operation mode:0 is c

Página 58

0x40 Set one of the step's current and time valuesByte offsetMeaning3 to 4 2 byte little-endian integer specifying which step number in the lis

Página 59

0x45 Set one of the step's power and time valuesByte offsetMeaning3 to 4 2 byte little-endian integer specifying which step number in the list5

Página 60

Byte offsetMeaning3 Partition scheme:1 means 1 file of 1000 list steps2 means 2 files of 500 list steps4 means 4 files of 250 list steps8 means 8 file

Página 61

Byte offsetMeaning7-24 Reserved0x50 Set timer value of for LOAD ONByte offsetMeaning3 to 4 2 byte little-endian integer specifying the time in units

Página 62

Byte offsetMeaning1 means to enable the Local key on the front panel4-24 Reserved0x56 Enable/disable remote sensingByte offsetMeaning3 0 means to di

Página 63 - 0x4D Recall the list file

Byte offsetMeaning3 Storage register, a number between 1 and 25 inclusive4-24 Reserved0x5C Recall DC Load's settingsByte offsetMeaning3 Storage

Página 64

The operation state register's bit meanings are:Bit Meaning0 Calculate the new demarcation coefficient1 Waiting for a trigger signal2 Remote cont

Página 65

Byte offsetMeaning3 Calibration protection state:Bit 0: value of 0 means the calibration state is not protectedBit 0: value of 1 means the calibrati

Página 66

The stored calibration data will be used the next time the DC Load is power cycled.0x67 Set calibration informationByte offsetMeaning3 to 22 ASCII i

Página 67

whose resistance doesn't change as a function of current or voltage).The DC Load can present dynamically changing loads to the DC source with mil

Página 68

Byte offsetMeaning12-24 Reserved8500 DC Load Series Version: November 4, 2008 Page 70 of 76

Página 69

Serial number and firmware versionTo find out the serial number and firmware version of the DC load, turn the instrument on. While the SYSTEM SELFTES

Página 70 - 12-24 Reserved

In case of troubleInstrument won't turn onIf the instrument won't turn on when the POWER switch is pressed in, please ensure the power cord

Página 71

Enter again. Now the password is cleared. Press Esc to exit the menu To unlock the other DC load models, use the following unlock keys:Model number

Página 72 - In case of trouble

Appendix: Service and warranty information SERVICE INFORMATION Warranty Service: Please go to the service and support section on our website www.bkp

Página 73

IndexAlphabetical IndexBattery...21Battery test...

Página 74

22820 Savi Ranch Parkway Yorba Linda CA, 92831Printed in China8500 DC Load Series Version: November 4, 2008 Page 76 of 76

Página 75 - Alphabetical Index

6123 4 5Button Function(s)116-character display shows voltage and current measurements.2Rotary knob. Turn to adjust a setting value. Press in to to

Página 76 - Printed in China

Standard displayThe standard display for the instrument is the voltage and current at the instrument's terminals. You may press the △ or ▽ keys

Comentários a estes Manuais

Sem comentários